Diagnostic Kit Instructions
Press the Suction
Pad against the
proximal surface of
the distal adapter
with the dome side
facing down.
Activate the vacuum.
Have the patient don the prosthesis:
fold down the top edge of the Alpha Liner
use a latex sleeve from the kit instead of
the Alpha Flex.
Activate the vacuum.
Press the Suction
Pad against the
proximal surface of
the distal adapter
with the dome side
facing up.
Activate the vacuum.
The leak
was coming
through the
socket walls.
Laminate a
new socket
with an inner
layer of
PETG.
To conrm that you have xed the leak:
Reassemble the pump and gasket to the socket.
Activate the vacuum.
Make sure that there are two
Proximal Gaskets in place
between the pump and the
themoplastic socket.
Make sure that the four M6
screws are tightened to 9
ft-lbs or 12 Nm.
Make sure that the internal
surface of the thermoplastic
socket is at and free of
defects.
The leak was
coming from the
Vacuum Pump.
Contact Ohio
Willow Wood.
Follow this process to evaluate the integrity of the
LimbLogic VS prior totting or to isolate the source
of a leak in the system.
The Alpha
Flex was the
source of the
leak.
Start HERE if the
leak is found AFTER
the prosthesis has
been delivered.
Start HERE if the
leak is found BEFORE
the prosthesis has
been delivered.
If the leak does NOT go away
For a LAMINATED
socket
For a THERMOPLASTIC
socket
If the leak does
NOT go away
If the leak
does NOT
go away
If the leak does
NOT go away
If the leak
goes away
If the leak goes away
If the leak
goes away
Renove the pump from the socket. Use the Diagnostic
Disc that is included with every pump to see if the leak is
coming from the pump:
Remove the protective tape from the Diagnostic Disc.
Place the Proximal Gasket in the recess on top of the
pump.
Place the Diagnostic Disc on top of the Proximal
Gasket.
Attach the pump with the Diagnostic Disc and
Proximal Gasket to the socket.
Start the pump and make sure the vacuum does not
bleed off.
The leak is coming from where the socket meets the pump. Possible causes:
The 4-Hole Distal Adapter was not attached to the socket using Fabtech 60 structural
urethane.
The pump was installed with the wrong bolts. If the 55 mm bolts for the thermoplastic
adapter were used in a laminated socket, the bolts will bottom out before they can be
appropriately tightened.
The leak could be coming from the proximal sealing gasket. Make sure the gasket is the
one that was provided by Ohio Willow Wood. Clean with isopropyl alcohol to remove
any debris. If the gasket is torn, install a new one.
The surface of the 4-Hole Distal Adapter and/or pump are damaged or dirty.
The walls of the center hole could be leaking. Drill the hole out to 3/8", ll the hole
with Fabtech 60 structural urethane, then drill out a new 9/32" hole.